Search
Close this search box.

Abstract

Aim: Develop and validate a volumetric absorptive microsampling (VAMS)-based LC–MS/MS method to support the bioanalysis of amino acid and carboxylic acid biomarkers in mouse whole blood. Method: Mouse whole blood was collected using a 10 μl VAMS device. The analytes in VAMS were extracted and analyzed using an LC–MS/MS method. Results: The VAMS-based LC–MS/MS assay exhibited a linearity range of 10.0–10,000 ng/ml with acceptable precision and accuracy and consistent recovery. The analyte stability in mouse whole blood VAMS was shown for 7 days at ambient conditions and at -80°C, as well as with three freeze/thaw cycles. Conclusion: A simple and robust VAMS-based LC–MS/MS method was developed and further validated for simultaneous bioanalysis of nine biomarkers in mouse whole blood.

Keywords:

  • amino acids
  • biomarkers
  • carboxylic acids
  • LC–MS/MS
  • microsampling
  • VAMS
Scroll to Top