Search
Close this search box.

Abstract

Aim: Microflow tandem mass spectrometry-based methods have been proposed as options to improve sensitivity and selectivity while improving sample utility and solvent consumption. Here, we evaluate a newly introduced microflow source, OptiFlow™, for quantitative performance. Results/methodology: We performed a comparison of the OptiFlow and IonDrive™ sources, respectively, on the same triple quadrupole mass spectrometer. The comparison used a neat cocktail of commercially available drugs and extracted plasma samples monitoring midazolam and alprazolam metabolites. Microflow produced a 2–4× signal increase for the neat drug cocktail and a 5–10× increase for extracted plasma samples. Conclusion: The OptiFlow method consistently gave increased signal response relative to the IonDrive method and enabled a better lower limit of quantitation for defining phamacokinetics.

Keywords:

  • LC–MS/MS
  • microflow
  • microsampling
  • pharmacokinetics