Aim: Ramucirumab, an anti-VEGFR2 monoclonal antibody, has been approved for the treatment of metastatic gastric and colorectal cancer. An assay measuring ramucirumab serum concentrations was needed to investigate its pharmacokinetics and concentration–response relationship. Results: An ELISA was developed and validated according to the international guidelines for ligand-binding assays. Ramucirumab calibration standards ranged from 0.125 to 40 mg/l. Low, middle and high quality controls were spiked at 0.2, 4 and 8 mg/l, respectively. The limits of quantification were established to be 0.125 and 10 mg/l for LLOQ and ULOQ, respectively. No cross-reactivity with anti-VEGF or anti-EGFR was detected. Conclusion: This in-house-developed ELISA is sensitive, accurate, reproducible and suitable for pharmacokinetic studies of ramucirumab.
Keywords:
- enzyme-linked immunosorbent assay
- monoclonal antibody
- pharmacokinetics
- ramucirumab
- therapeutic drug monitoring