Search
Close this search box.

Abstract

Background: Many bioanalytical methods for antisense oligonucleotides (ASOs) using LC–MS have been reported. However, no data have been available on the reproducibility and robustness of a single bioanalytical method for ASOs. As such, in the current study, we evaluated the reproducibility and robustness of LC–MS-based bioanalytical methods for ASOs in multiple laboratories. Methods/Results: Seven independent laboratories were included in this study. Mipomersen was measured by ion-pairing LC–MS (IP-LC–MS) as a model ASO using different LC–MS. The validation results of calibration curve, accuracy, precision and selectivity met the criteria of conventional bioanalytical method validation guidelines using LC/GC–MS for drugs in all laboratories. Meanwhile, carryover (>20%) was detected in three laboratories. Conclusion: We first demonstrated the multicenter-validated IP-LC–MS bioanalytical method for ASOs. Our data showed that the method was sensitive, robust and reproducible. However, the occurrence of carryover should be carefully monitored in its future application.

Keywords:

  • antisense oligonucleotide
  • Clarity OTX
  • ion-pairing reagent
  • LC–MS bioanalysis
  • multilaboratory validation
Scroll to Top