Abstract

Aim: A novel HPLC method was developed and validated for the simultaneous estimation of paclitaxel (PTX) and baicalein (BAC). Materials & methods: The analytes were resolved in a C18 column using the aqueous solution of formic acid (0.10% v/v) and MeOH (30:70 v/v). Results: The developed method was found to be linear over the concentration ranges 0.039–10 μg/ml and 0.019–10 μg/ml for PTX and BAC, respectively. The lower limits of quantification obtained were 0.042 μg/ml and 0.361 μg/ml for PTX and BAC, respectively. Conclusion: The developed method was found to be precise and accurate as per the International Council for Harmonisation of Technical Requirements for Pharmaceuticals for Human Use guidelines, for simultaneous estimation of PTX and BAC, having an application in formulation development and bioanalytical studies.

Keywords:

  • baicalein
  • bioanalytical method
  • HPLC
  • paclitaxel
  • simultaneous quantification